We have revisited the NGC 6633 calibration images used in the early MLO Earthshine reductions. This corrects the interpretation in Filter Wheel and/or Shutter annoy, again, and clarifies the numerical labelling on Instrumental zero points. It also checks the filter assignments used in PSF profiles from the NGC6633 images.
Two filter-wheel failures
Catalogue photometry of the stars identifies two 2 s exposures recorded as IRCUT whose physical passband was VE1:
| Julian Date | Recorded | Inferred | Exposure | Probability |
|---|---|---|---|---|
| 2456104.9536571 | IRCUT | VE1 | 2 s | 0.942 |
| 2456104.9626903 | IRCUT | VE1 | 2 s | 0.944 |
The second exposure is the third of four usable exposures in its IRCUT sequence. The 2012 recommendation to discard the first and last exposure of each sequence therefore does not remove every wheel failure. The statement that the NGC 6633 images showed an absence of the filter-wheel problem was incorrect.
The published calibration sequence is secure
The zero-point analysis used 25 s exposures in V, 34 s in B, 12 s in VE1, 32 s in VE2 and 12 s in IRCUT. We find 68 usable images with exactly these settings. Their stellar colours confirm every recorded filter assignment:
| Filter | Usable images | Confirmed | Ambiguous | Mismatch |
|---|---|---|---|---|
| B | 20 | 20 | 0 | 0 |
| V | 8 | 8 | 0 | 0 |
| VE1 | 12 | 12 | 0 | 0 |
| VE2 | 16 | 16 | 0 | 0 |
| IRCUT | 12 | 12 | 0 | 0 |
The two failed 2 s IRCUT exposures precede this long-exposure sequence and were not used in the published calibration. The five-filter calibration therefore used the intended physical filters. The later B−V Earthshine results are unchanged. The 2013 point-spread-function post states that it used the same calibration images, so its B, V, VE1, VE2 and IRCUT assignments are also supported; the page does not list the individual raw filenames.
Corrected calibration equations
For the archive-wide fit to dark-corrected images without flat fields, instrumental magnitude is minst = −2.5 log10(ADU s−1) and the extinction-corrected magnitude is m0 = minst − kX. After reassigning the two failed exposures, the catalogue transformations are:
NGC 6633
B = b0 + 15.038 + 0.152(B−V)
V = v0 + 15.233 − 0.063(B−V)
V = ve10 + 16.661 + 0.112(B−V)
V = ve20 + 14.309 + 1.082(B−V)
V = ircut0 + 16.526 + 0.137(B−V)
The fitted extinction coefficients are retained separately as part of the atmospheric reduction; they are not coefficients of the catalogue transformation.
Sensitivity of an archive-wide recalibration
As a separate sensitivity test, we fitted the transformations to every NGC 6633 exposure with usable catalogue photometry, first with the recorded labels and then with the two exposures reassigned from IRCUT to VE1. These are the changes for the dark-corrected reduction without flat fields:
| Transformation | Change in zero point at unit airmass (mag) | Change in colour coefficient | Largest zero-point change over observed airmass (mag) |
|---|---|---|---|
| B to B | 0 | 0 | 0 |
| V to V | 0 | 0 | 0 |
| VE1 to V | −0.00140 | −0.00015 | 0.00181 |
| VE2 to V | 0 | 0 | 0 |
| IRCUT to V | −0.00528 | +0.00167 | 0.00651 |
The B and V transformations are exactly unchanged. The VE1 and IRCUT changes remain below 0.007 mag over the observed airmass range. With flat-fielded images, the IRCUT fit was unchanged because the robust fit had already rejected the two discrepant exposures; the VE1 zero point changed by −0.00183 mag at unit airmass.
A second correction to the 2012 text
The instrumental-zero-point page contains an internal labelling inconsistency unrelated to the wheel failures. Its first equation block gives V and B zero points of 15.07 and 14.75. The following text assigns those values to M41 and gives 15.15 and 14.46 for NGC 6633. Future quantitative use will come from the frame-level recalibration with its stated aperture and extinction conventions.
Remaining uncertainty
VE1 and IRCUT have similar passbands, and the camera records the requested filter rather than independent wheel-position telemetry. The two assignments therefore rest on the empirical stellar colour locus. Their probabilities are 0.942 and 0.944. This limitation does not affect the confirmed long-exposure calibration sequence or its B−V results.